Data-dependent acquisition based high-resolution mass spectrum for trace Alternaria mycotoxin analysis and sulfated metabolites identification

Food Chem. 2021 Dec 1:364:130450. doi: 10.1016/j.foodchem.2021.130450. Epub 2021 Jun 24.

Abstract

Alternaria mycotoxins are food-related compounds that are mainly produced by Alternaria fungi species. However, it's difficult for Alternaria mycotoxins analysis, especially for conjugated metabolites in food safety surveillance. In this work, a novel data-dependent acquisition (DDA) full mass scan and products scan protocol was proposed for qualitative and quantitative analysis of five target mycotoxins in tomato samples using ultra-high-performance liquid chromatography coupled with quadrupole Orbitrap high-resolution mass spectrometry (UHPLC-Q-Orbitrap). In total, 24 sulfated metabolites were detected with post-data analysis techniques, and two sulfated metabolites (AME-sulfated and AOH-sulfated) were identified in Alternaria fungi -inoculated tomatoes. In addition, a custom database was established, and it was successfully applied for Alternaria mycotoxins and sulfated metabolites screening in tomatoes. With the improvement in high-resolution mass spectrometry (HRMS) as well as post-data analysis techniques, DDA based HRMS method could be widely applied for compound analysis, identification, and screening in quantitative field.

Keywords: Alternaria mycotoxin; Data-dependent acquisition; Food safety; High-resolution mass spectrometer; Sulfated metabolites.

MeSH terms

  • Alternaria*
  • Chromatography, High Pressure Liquid
  • Food Contamination / analysis
  • Mass Spectrometry
  • Mycotoxins* / analysis
  • Sulfates

Substances

  • Mycotoxins
  • Sulfates