[Usefulness of Combining Post-processing Scatter Correction and an Anti-scatter Grid in Chest Standing Radiography]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2021;77(6):555-563. doi: 10.6009/jjrt.2021_JSRT_77.6.555.
[Article in Japanese]

Abstract

Purpose: The aim of this study was to evaluate the usefulness of combining post-processing scatter correction (IG) and an anti-scatter grid (RG) in chest radiography.

Method: To determine the combination protocol (Hyb) that was closed to RG 12:1 (RG12), we measured the content rate of scattered radiation for each combination (RG12, IG12, RG3-12+IG3-12). Task-based modulation transfer function (MTF_Task) and SDNR were evaluated using RG12, IG12, and Hyb. Additionally, seven radiologists performed visual evaluation by using chest phantom.

Result: The protocol of Hyb was RG8+IG3. In SDNR, Hyb (RG8+IG3) was equal to or higher than RG12, and MTF_Task was equal in all grid systems. Hyb (RG8+IG3) was significantly superior to RG12 in visual evaluation.

Conclusion: The combining post-processing scatter correction should be useful for improving inspection throughput and reducing the risk of grid's damage.

Keywords: anti-scatter grid; chest radiography; flat panel detector (FPD); image processing; scatter correction.