Pediatric reference interval verification for common biochemical assays on the Abbott Alinity system

Clin Chem Lab Med. 2021 May 19;59(9):1554-1562. doi: 10.1515/cclm-2021-0336. Print 2021 Aug 26.

Abstract

Objectives: The quality of clinical laboratory service depends on quality laboratory operations and accurate test result interpretation based on reference intervals (RIs). As new analytical systems continue to be developed and improved, previously established RIs must be verified. The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) has established comprehensive RIs for many biomarkers on several analytical systems. Here, published CALIPER RIs for 28 chemistry assays on the Abbott ARCHITECT were assessed for verification on the newer Alinity system.

Methods: An analytical validation was first completed to assess assay performance. CALIPER serum samples (100) were analyzed for 28 chemistry assays on the Alinity system. The percentage of results falling within published pediatric ARCHITECT reference and confidence limits was determined for each analyte. Based on Clinical and Laboratory Standards Institute (CLSI) guidelines, if ≥90% of test results fell within confidence limits of ARCHITECT assay RIs, they were considered verified.

Results: Of the 28 assays assessed, 26 met the criteria for verification. Reference values for calcium and magnesium did not meet the criteria for verification with 87% and 35% falling within previously established ARCHITECT confidence limits, respectively. However, both assays could be verified using pediatric RIs provided in the Abbott Alinity package insert.

Conclusions: In this study, CALIPER ARCHITECT RIs were verified on the Alinity system for several chemistry assays. These data demonstrate excellent concordance for most assays between the Abbott ARCHITECT and Alinity systems and will assist in the implementation of the Alinity system in pediatric healthcare institutions.

Keywords: Abbott Alinity; CALIPER; chemistry; pediatric; reference intervals.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biological Assay* / methods
  • Child
  • Clinical Laboratory Services*
  • Humans
  • Magnesium
  • Reference Standards
  • Reference Values

Substances

  • Magnesium