Background: The purpose of this study was to determine functional outcomes, implant survival rate, and complications of a single-radius, high-flexion posterior stabilized (PS) total knee prosthesis at a minimum follow up of 10 years for a consecutive series.
Methods: This retrospective observational study included 395 consecutive patients who underwent 585 single-radius, PS total knee arthroplasties (TKAs) between January 2009 and December 2009. Their functional outcomes, implant survival rates, radiological findings, and complications were evaluated.
Results: At a mean follow up of 11.2 years (range, 10.6-11.8 years), 395 (80.8%) patients were available for review. At final follow up, preoperative Knee Society knee scores were improved from 37.3 to 80.1 (P < 0.05) and function scores were improved from 35.7 to 80.5 points (P < 0.05). Sixteen knees (2.7%) in 15 patients required revision surgery due to 11 septic loosening (1.9%) and five aseptic loosening events (0.8%). Cumulative survival for the prosthesis was 97.3% for any cause at 10 years. A total of 34 radiolucent lines (5.8%) were detected an average of 2.4 years following surgery and radiological assessment did not reveal any evidence of component migration at final follow up.
Conclusions: Single-radius, high-flexion PS TKA showed good long-term survival rates and clinical outcomes. Further study is required to determine whether limited radiolucency findings at the tibial component-tibial plateau could progress to or result in the loosening of components.
Keywords: Outcome; Posterior stabilized; Single radius; Survival; Total knee arthroplasty.