Radiocarbon dioxide detection using cantilever-enhanced photoacoustic spectroscopy

Opt Lett. 2021 May 1;46(9):2083-2086. doi: 10.1364/OL.420199.

Abstract

In this Letter, we report on the sub-parts-per-billion-level radiocarbon dioxide detection using cantilever-enhanced photoacoustic spectroscopy. The 14C/C ratio of samples is measured by targeting a 14CO2 absorption line with minimal interference from other CO2 isotopes. Using a quantum cascade laser as a light source allows for a compact experimental setup. In addition, measurements of sample gases with 14CO2 concentrations as low as 100 parts-per-trillion (ppt) are presented. The Allan deviation demonstrates a noise equivalent concentration of 30 ppt at an averaging time of 9 min. The achieved sensitivity validates this method as a suitable alternative to more complex optical detection methods for radiocarbon dioxide detection used so far, and it can be envisioned for future in situ radiocarbon detection.