Use of the FebriDx point-of-care test for the exclusion of SARS-CoV-2 diagnosis in a population with acute respiratory infection during the second (COVID-19) wave in Italy

Int J Infect Dis. 2021 Jul:108:231-236. doi: 10.1016/j.ijid.2021.04.065. Epub 2021 Apr 24.

Abstract

Objective: Evaluate the real-world accuracy of Myxovirus resistance protein A (MxA) detected by the rapid, point-of-care FebriDx test during the second-wave pandemic in Italy in patients with acute respiratory infection (ARI) and a clinical suspicion of COVID-19.

Design and methods: Prospective, observational, diagnostic accuracy study whereby hospitalized patients with ARI were consecutively enrolled in a single tertiary care center in Italy from August 1, 2020 to January 31, 2021.

Results: COVID-19 was diagnosed in 136/200 (68.0%) patients and Non-COVID-19 was diagnosed in 64/200 (32.0%) patients. COVID-19 patients were younger and had a lower Charlson comorbidity index compared to Non-COVID-19 patients (p < 0.001). Concordance between FebriDx, MxA and rt-PCR for SARS-CoV-2 (gold standard) was good (k 0.93, 95% CI 0.87-0.99). Overall sensitivity and specificity were 97.8% [95% CI 93.7-99.5] and 95.3% [95% CI 86.9%-99.0%], respectively. FebriDx demonstrated a negative predictive value of 95.3% (95% CI 86.9-99.0) for an observed disease prevalence of 68%.

Conclusions: FebriDx MxA showed high diagnostic accuracy to identify COVID-19 and could be considered as a real-time triage tool to streamline the management of suspected COVID-19 patients. FebriDx also detected bacterial etiology in Non-COVID-19 patients suggesting good performance to distinguish bacterial from viral respiratory infection.

Keywords: Accuracy; COVID-19; FebriDx; Italy; Point-of-care.

MeSH terms

  • COVID-19 Testing
  • COVID-19*
  • Humans
  • Italy / epidemiology
  • Point-of-Care Testing
  • Prospective Studies
  • SARS-CoV-2*
  • Sensitivity and Specificity