Evaluation of latest viscoelastic coagulation assays in the transcatheter aortic valve implantation setting

Open Heart. 2021 Apr;8(1):e001565. doi: 10.1136/openhrt-2020-001565.

Abstract

Background: Point of care viscoelastic measures with thromboelastography (TEG; Haemonetics Corporation, Switzerland) and thromboelastometry (ROTEM, Tem Innovations GmbH, Germany) now supersede laboratory assays in the perioperative assessment and management of coagulation. To the best of our knowledge, this sophisticated coagulation assessment has not been performed to characterise thrombotic changes in the transcatheter aortic valve implantation (TAVI) setting, nor have the two latest iteration cartridge-based systems been directly compared in the elective perioperative period.

Methods: Patients undergoing TAVI were prospectively recruited. Samples (n=44) were obtained at four timepoints (postinduction of anaesthesia, postheparin (100 IU/kg), postprotamine (1 mg/100 IU heparin) and 6 hours postoperatively). Each sample was concurrently assessed with standard laboratory tests (prothrombin time/international normalised ratio, activated partial thromboplastin time, thrombin clotting time, platelet count and direct fibrinogen, ROTEMSigma and TEG6s).

Results: Clot strength showed a statistically significant increase postheparin/TAVI deployment. When considering the subgroup of samples taken following the administration heparin, the heparinase channel of the TEG6s did not yield clotting strength results in 55% of samples and clotting time exceeded the upper limit of normal in 70% of samples. It was retrospectively recognised that the arachidonic acid channel of the TEG6s Platelet Mapping Cartridge had been decommissioned prohibiting assessment of aspirin effect.

Conclusions: This study demonstrated a small intraprocedural prothrombotic change of uncertain clinical importance during the transcatheter aortic valve procedure. Further comparison with percutaneous coronary intervention and aortic valve replacement cohorts are needed to assess the merits of current antithrombotic guidelines, which are extrapolated from the PCI setting. The heparin effect was more consistently quantified by ROTEM.

Keywords: aortic valve stenosis; drug monitoring; heart valve prosthesis; transcatheter aortic valve replacement.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged, 80 and over
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / blood
  • Aortic Valve Stenosis / surgery
  • Blood Coagulation / physiology*
  • Blood Coagulation Tests / methods*
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Prospective Studies
  • Thrombelastography / methods
  • Transcatheter Aortic Valve Replacement / methods*