A Dual-Platform Point-of-Care Test for Neurosyphilis Diagnosis

Sex Transm Dis. 2021 May 1;48(5):353-356. doi: 10.1097/OLQ.0000000000001308.

Abstract

Background: The diagnosis of neurosyphilis relies on cerebrospinal fluid (CSF) abnormalities (pleocytosis, elevated protein) and CSF-Venereal Disease Research Laboratory (VDRL) test. In resource-limited settings, the CSF-VDRL test may not be widely available.

Methods: We optimized a commercial immunochromatographic strip test, the DPP Chembio syphilis assay, for performance with CSF and tested centrifuged CSF samples of 71 patients with syphilis (35 with neurosyphilis and 36 without neurosyphilis). A CSF dilution of 1:4 was chosen based on agreement with CSF pools with documented results from the CSF-VDRL test and fluorescent treponemal antibody absorption test on CSF. Using an electronic reader, we obtained unit values of treponemal and nontreponemal antibodies for all study samples and generated a receiver operating characteristic curve; using the Youden index, we established diagnostic cutoffs with optimal sensitivity and specificity.

Results: Diagnostic sensitivity of the nontreponemal test was 80% (95% confidence interval, 63%-92%) and specificity was 97% (95% confidence interval, 85%-100%) for neurosyphilis diagnosis using a reactive CSF-VDRL that improved after neurosyphilis therapy as a criterion standard.

Conclusions: In this small study, the DPP Chembio test showed promising results for neurosyphilis diagnosis. Further studies are needed to assess its performance in resource-limited settings.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Fluorescent Treponemal Antibody-Absorption Test
  • Humans
  • Neurosyphilis* / diagnosis
  • Point-of-Care Testing
  • Syphilis Serodiagnosis
  • Treponema pallidum*