TOPAS MC software was used to model the efficiency of a coaxial p-type HPGe detector, type GX9023 from Canberra. The model was validated by comparing experimental efficiencies with efficiencies calculated by TOPAS MC simulations. Three different geometries of radionuclide sources, placed at different heights from the detector endcap, were used to validate the model. The imposed criteria of 5% relative difference was met for a range of radionuclides and gamma-ray energies. As a result, the created detector model with TOPAS MC was considered validated.
Keywords: High purity germanium detector; Monte Carlo; TOPAS MC.
Copyright © 2021 Elsevier Ltd. All rights reserved.