Digital adaptive optics based on digital lateral shearing of the computed pupil field for point scanning retinal swept source OCT

Biomed Opt Express. 2021 Feb 23;12(3):1577-1592. doi: 10.1364/BOE.416569. eCollection 2021 Mar 1.

Abstract

A novel non-iterative digital adaptive optics technique is presented in which the wavefront error is calculated using the phase difference between the pupil field and its digital copies translated by a pixel along the horizontal and vertical direction in the pupil plane. This method provides slope data per pixel, thus can generate > 50k local slope data samples for a circular pupil of diameter 256 pixels with high accuracy and dynamic range. It offers more than 12x faster computational speed in comparison to the sub-aperture based digital adaptive optics method. Furthermore, it is independent of any system parameters, the light distribution in the pupil plane, or the intensity of the image. The technique is useful in applications such as interferometric or digital holography based microscopy, metrology, and as digital wavefront sensor in adaptive optics, where focusing of light in the sample is involved that creates a guide star or where the sample itself exhibits guide star-like structures. This technique is implemented in a point scanning swept-source OCT at 1060 nm, and a large wavefront error with a peak to valley of 20 radians and root mean square error of 0.71 waves is detected and corrected in case of a micro-beads phantom sample. Also, human photoreceptor images are recovered from aberrated retinal OCT volumes acquired at eccentricities of 2 and 2.5 degrees from the fovea in vivo.