A method for quantitative evaluations of scanning-proton dose distributions

J Appl Clin Med Phys. 2021 Apr;22(4):193-201. doi: 10.1002/acm2.13226. Epub 2021 Mar 29.

Abstract

Purpose: Patient-Specific Quality Assurance (PSQA) measurement analysis depends on generating metrics representative of calculation and measurement agreement. Considering the heightened capability of discrete spot scanning protons to modulate individual dose voxels, a dose plane comparison approach that maintained all of the capabilities of the well-established γ test, but that also provided a more intuitive error parameterization, was desired.

Methods: Analysis was performed for 300 dose planes compared by searching all calculated points within a fixed radius around each measured pixel to determine the dose deviation. Dose plane agreement is reported as the dose difference minimum (DDM) within an empirically established search radius: ΔDmin(r). This per-pixel metric is aggregated into a histogram binned by dose deviation. Search-radius criteria were based on a weighted-beamlet 3σ spatial deviation from imaging isocenter. Equipment setup error was mitigated during analysis using tracked image registration, ensuring beamlet deviations to be the dominant source of spatial error. The percentage of comparison points with <3% dose difference determined pass rate.

Results: The mean beamlet radial deviation was 0.38mm from x-ray isocenter, with a standard deviation of 0.19mm, such that 99.9% of relevant pencil beams were within 1 mm of nominal. The dose-plane comparison data showed no change in passing rate between a 3%/1mm ΔDmin(r) analysis (97.6 +/- 3.6%) and a 3%/2mm γ test (97.7 +/- 3.2%).

Conclusions: PSQA dose-comparison agreements corresponding to a search radius outside of machine performance limits are likely false positives. However, the elliptical shape of the γ test is too dose-restrictive with a spatial-error threshold set at 1 mm. This work introduces a cylindrical search shape, proposed herein as more relevant to plan quality, as part of the new DDM planar-dose comparison algorithm. DDM accepts all pixels within a given dose threshold inside the search radius, and carries forward plan-quality metrics in a straightforward manner for evaluation.

Keywords: PSQA; gamma test; proton.