Validation of the Microlife BP B3 AFIB upper arm blood pressure monitor in adults and adolescents according to the ANSI/AAMI/ISO 81060-2:2019 protocol

Blood Press Monit. 2021 Aug 1;26(4):299-304. doi: 10.1097/MBP.0000000000000530.

Abstract

Objective: Aim of this study was to validate the Microlife BP B3 AFIB/enterprise resource planning (ERP) No: BP3KT1-3 N blood pressure (BP) monitor according to the American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation (AAMI)/International Organization for Standardization (ISO) 81060-2:2019 in adolescents and adults from a general population.

Methods: BP measurements on the upper arm were performed in 85 subjects (age range 12-88 years), using the Microlife BP B3 AFIB and a standard mercury reference sphygmomanometer.

Results: A total of 255 valid BP comparisons were performed for the present validation analysis. The mean ± SD difference between the test and the reference device was 0.70 ± 7.05 mmHg for SBP (pass criterion ≤5 mmHg) and -0.85 ± 4.70 mmHg for DBP (pass criterion ≤5 mmHg) with the SD below the required value of ≤8 mmHg. The mean ± SD of the intraindividual differences between the test and the reference device was 0.70 ± 5.87 mmHg for SBP (pass criterion for the SD ≤6.90 mmHg) and -0.85 ± 4.19 mmHg for DBP (pass criterion for the SD ≤6.88 mmHg).

Conclusion: The Microlife BP B3 AFIB/ERP No: BP3KT1-3 N has passed the criteria of the ANSI/AAMI/ISO 81060-2:2019 protocol and can be recommended for home BP measurements in adolescents and adults.

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Arm
  • Atrial Fibrillation*
  • Blood Pressure
  • Blood Pressure Determination
  • Blood Pressure Monitors*
  • Child
  • Humans
  • Middle Aged
  • Reference Standards
  • Sphygmomanometers
  • Young Adult