Long bone fracture reduction and deformity correction using the hexapod external fixator with a new method: a feasible study and preliminary results

BMC Musculoskelet Disord. 2021 Feb 24;22(1):221. doi: 10.1186/s12891-021-04097-9.

Abstract

Background: The hexapod external fixator (HEF), such as the Taylor spatial frame (TSF), offering the ability of multidirectional deformities correction without changing the structure, whereas there are so many parameters for surgeons to measure and subjective errors will occur inevitably. The purpose of this study was to evaluate the effectiveness of a new method based on computer-assisted three-dimensional (3D) reconstruction and hexapod external fixator for long bone fracture reduction and deformity correction without calculating the parameters needed by the traditional usage.

Methods: This retrospective study consists of 25 patients with high-energy tibial diaphyseal fractures treated by the HEF at our institution from January 2016 to June 2018, including 22 males and 3 females with a mean age of 42 years (range 14-63 years). Hexapod external fixator treatments were conducted to manage the multiplanar posttraumatic deformities with/without poor soft-tissue that were not suitable for internal fixation. Computer-assisted 3D reconstruction and trajectory planning of the reduction by Mimics were applied to perform virtual fracture reduction and deformity correction. The electronic prescription derived from the length changes of the six struts were calculated by SolidWorks. Fracture reduction was conducted by adjusting the lengths of the six struts according to the electronic prescription. Effectiveness was evaluated by the standard anteroposterior (AP) and lateral X-rays after reduction.

Results: All patients acquired excellent functional reduction and achieved bone union in our study. After correction, the mean translation (1.0 ± 1.1 mm) and angulation (0.8 ± 1.2°) on the coronal plane, mean translation (0.8 ± 1.0 mm) and angulation (0.3 ± 0.8°) on the sagittal plane were all less than those (6.1 ± 4.9 mm, 5.2 ± 3.2°, 4.2 ± 3.5 mm, 4.0 ± 2.5°) before correction (P < 0.05).

Conclusions: The computer-assisted three-dimensional reconstruction and hexapod external fixator-based method allows surgeons to conduct long bone fracture reduction and deformity correction without calculating the parameters needed by the traditional usage. This method is suggested to apply in those unusually complex cases with extensive soft tissue damage and where internal fixation is impossible or inadvisable.

Keywords: Computer‐assisted; Fracture reduction; Hexapod external fixation; Taylor spatial frame; Three‐dimensional reconstruction.

MeSH terms

  • Adolescent
  • Adult
  • External Fixators*
  • Female
  • Fracture Fixation
  • Humans
  • Male
  • Middle Aged
  • Retrospective Studies
  • Tibia
  • Tibial Fractures*
  • Young Adult