Self-Limiting Mussel Inspired Thin Antifouling Coating with Broad-Spectrum Resistance to Biofilm Formation to Prevent Catheter-Associated Infection in Mouse and Porcine Models

Adv Healthc Mater. 2021 Mar;10(6):e2001573. doi: 10.1002/adhm.202001573. Epub 2021 Jan 20.

Abstract

Catheter-associated urinary tract infections (CAUTIs) are one of the most commonly occurring hospital-acquired infections. Current coating strategies to prevent catheter-associated biofilm formation are limited by their poor long-term efficiency and limited applicability to diverse materials. Here, the authors report a highly effective non-fouling coating with long-term biofilm prevention activity and is applicable to diverse catheters. The thin coating is lubricous, stable, highly uniform, and shows broad spectrum prevention of biofilm formation of nine different bacterial strains and prevents the migration of bacteria on catheter surface. The coating method is adapted to human-sized catheters (both intraluminal and extraluminal) and demonstrates long-term biofilm prevention activity over 30 days in challenging conditions. The coated catheters are tested in a mouse CAUTI model and demonstrate high efficiency in preventing bacterial colonization of both Gram-positive and Gram-negative bacteria. Furthermore, the coated human-sized Foley catheters are evaluated in a porcine CAUTI model and show consistent efficiency in reducing biofilm formation by Escherichia coli (E. coli) over 95%. The simplicity of the coating method, the ability to apply this coating on diverse materials, and the high efficiency in preventing bacterial adhesion increase the potential of this method for the development of next generation infection resistant medical devices.

Keywords: animal models of infection; antifouling coating; catheter-associated urinary tract infection; infection resistant surfaces; long-term biofilm prevention activity.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Anti-Bacterial Agents
  • Biofilms
  • Catheter-Related Infections* / prevention & control
  • Escherichia coli
  • Gram-Negative Bacteria
  • Gram-Positive Bacteria
  • Mice
  • Swine
  • Urinary Catheters

Substances

  • Anti-Bacterial Agents

Grants and funding