Evaluation of BD Phoenix and Microscan WalkAway for determination of fosfomycin susceptibility in Staphylococcus aureus

Diagn Microbiol Infect Dis. 2021 Apr;99(4):115240. doi: 10.1016/j.diagmicrobio.2020.115240. Epub 2020 Oct 17.

Abstract

We evaluate the performance of BD Phoenix (BD Diagnostic; Hunt Valley, MD) and MiscroScan WalkAway (Beckman Coulter Inc; Carlsbad, CA) systems versus reference agar dilution method for fosfomycin susceptibility determination in 200 Staphylococcus aureus clinical isolates. Both methods exhibit ≤89.0% of categorica agreement and ≥63.3% of very major error. All commercial antimicrobial susceptibility systems show poor concordance with reference method.

Keywords: Agar Dilution; Fosfomycin; Methicillin-Resistant Staphylococcus aureus; Staphylococcus aureus.

MeSH terms

  • Anti-Bacterial Agents / pharmacology*
  • Bacteriological Techniques*
  • Drug Resistance, Bacterial*
  • Fosfomycin / pharmacology*
  • Staphylococcus aureus / drug effects*

Substances

  • Anti-Bacterial Agents
  • Fosfomycin