Concurrent Validity and Reliability of a Low-Cost Dynamometer to Assess Maximal Isometric Strength in Neck Movements

J Manipulative Physiol Ther. 2021 Mar;44(3):229-235. doi: 10.1016/j.jmpt.2020.08.005. Epub 2021 Jan 15.

Abstract

Objective: This study aimed to evaluate the concurrent validity and reliability of a low-cost digital dynamometer to assess maximal isometric strength in neck movements.

Methods: Twelve recreationally active participants (6 women, 6 men; age: 24.1 ± 3.4 years; height: 1.71 ± 0.06 m; weight: 64.6 ± 11.7 kg) performed neck flexion, extension, and lateral flexion, and maximal isometric strength was simultaneously measured using an isokinetic dynamometer (gold standard) and a digital dynamometer to evaluate the concurrent validity of the latter. One week later, all the participants repeated 3 trials of each neck movement (3 minutes apart) registered only with the digital dynamometer. The first and second trial were guided by tester 1 and the third trial was guided by tester 2.

Results: The concurrent validity of the digital dynamometer showed a nearly perfect correlation between both devices (r > 0.986, P < .001), with Bland-Altman plots showing absolute agreement. Intertester and intratester reliability were nearly perfect for all neck movements (intraclass correlation coefficient > 0.86).

Conclusion: The low-cost digital dynamometer showed valid and reliable measurements of maximal isometric strength in neck movements.

Keywords: Isometric Contraction; Muscle Strength Dynamometer; Reproducibility of Results.

Publication types

  • Validation Study

MeSH terms

  • Adult
  • Female
  • Humans
  • Isometric Contraction / physiology*
  • Male
  • Muscle Strength / physiology*
  • Muscle Strength Dynamometer / standards*
  • Range of Motion, Articular / physiology*
  • Reproducibility of Results
  • Young Adult