[Validation of point-of-care device for rapid detection of urinary tract infection and antibiotic susceptibility]

Rev Chilena Infectol. 2020 Nov;37(5):523-530. doi: 10.4067/S0716-10182020000500523.
[Article in Spanish]

Abstract

Background: Urinary tract infections (UTI) presents a high prevalence in the community setting. Rapid and accurate microbiological diagnosis is essential to ensure adequate and effective therapy.

Aim: To evaluate a rapid antibiogram kit (KAR®) in point-of-care format for rapid detection of UTI and antibiotic susceptibility.

Methods: The KAR® device has been designed and developed in collaboration with technical engineers and clinical microbiologists. Its evaluation has been carried out through a multicenter study in which three Spanish hospitals have participated. Thus, different in vivo tests have been implemented in order to determine device correlation with the reference microbiological techniques.

Results: During the study period, a total of 400 urine samples from patients with suspected ITU were tested. The KAR® device provided fast results (mean positivity time of 7,8 ± 1,5 hours) with 97% sensitivity, 89% specificity and 87% agreement for the detection of significant bacteriuria. The percentages of specificity for the antibiotics tested were: ciprofloxacin (97%), fosfomycin (94%),cotrimoxazole (84%), ampicillin (80%) and amoxicillin/clavulanic acid (55%).

Conclusion: The KAR® device could be a useful tool for diagnosing UTI in outpatients, especially in areas of low socio-economic level.

Publication types

  • Multicenter Study
  • Validation Study

MeSH terms

  • Anti-Bacterial Agents / therapeutic use
  • Humans
  • Microbial Sensitivity Tests
  • Point-of-Care Systems*
  • Prevalence
  • Urinary Tract Infections* / diagnosis
  • Urinary Tract Infections* / drug therapy

Substances

  • Anti-Bacterial Agents