Portable fixed dynamometry: towards remote muscle strength measurements in patients with motor neuron disease

J Neurol. 2021 May;268(5):1738-1746. doi: 10.1007/s00415-020-10366-9. Epub 2020 Dec 23.

Abstract

Background: We aimed to determine (1) the test-retest reliability of a newly developed portable fixed dynamometer (PFD) as compared to the hand-held dynamometer (HHD) in patients with motor neuron disease (MND) and (2) the PFD's ability to reduce possible examiner-induced ceiling effects.

Methods: Test-retest reliability of isometric muscle strength of the quadriceps was measured in patients with MND and non-neurological controls using the HHD and PFD. Reliability was estimated by the intraclass correlation coefficient (ICC) and standard error of measurement (SEM) using linear mixed effects models, and the Bland-Altman method of agreement.

Results: In total, 45 patients with MND and 43 healthy controls were enrolled in this study. The ICC of the PFD was excellent and similar in both patients and controls (ICC Patients 99.5% vs. ICC Controls 98.6%) with a SEM of 6.2%. A strong examiner-induced ceiling effect in HHD was found when the participant's strength exceeded that of examiner. Employing the PFD increased the range of muscle strength measurements across individuals nearly twofold from 414 to 783 N.

Conclusions: Portable fixed dynamometry may significantly reduce examiner-induced ceiling effects, optimize the standardization of muscle strength testing, and maximize reliability. Ultimately, PFD may improve the delivery of care due to its potential for unsupervised, home-based assessments and reduce the burden to the patient of participating in clinical trials for MND or other neuromuscular diseases.

Keywords: Fixed dynamometry; Motor neuron disease; Muscle strength; Reliability study; Remote monitoring.

MeSH terms

  • Humans
  • Isometric Contraction*
  • Motor Neuron Disease*
  • Muscle Strength
  • Muscle Strength Dynamometer
  • Reproducibility of Results