Abstract

This study investigated the cyclic fatigue, bending, torsional resistance and angular deflection of Pro-R 25.08, Logic RT 25.08, MK Retreatment 25.08 (MK RT) and ProTaper Retreatment D2 instruments. Cyclic fatigue test was performed until fracture in a custom stainless-steel device with water bath equipment to simulate body temperature. Fracture time was recorded. A number of cycles were calculated. Resistance to bending at 45°, torsional resistance and angular deflection were evaluated. The fracture surfaces were examined by scanning electron microscopy. Statistical analysis was performed with one-way ANOVA and Tukey and Kruskal-Wallis and Dunn tests. Pro-R and Logic RT presented the highest cyclic fatigue (P < 0.05). D2 had higher strength to bend than Pro-R and Logic RT (P < 0.05). Logic RT showed the highest torque and angular deflection (P < 0.05), without differences when compared to D2 (P > 0.05). Instrument performances were dependent on their geometrical features and heat treatments.

Keywords: NiTi alloy; body temperature; cyclic fatigue; strength to bend; torsion.

MeSH terms

  • Dental Alloys
  • Equipment Design
  • Equipment Failure
  • Materials Testing
  • Nickel*
  • Retreatment
  • Root Canal Preparation
  • Stress, Mechanical
  • Titanium*
  • Torsion, Mechanical

Substances

  • Dental Alloys
  • Nickel
  • Titanium