Effect of preventing air intrusion on silent strokes during atrial fibrillation ablation using a mini-basket catheter

Pacing Clin Electrophysiol. 2021 Jan;44(1):71-81. doi: 10.1111/pace.14131. Epub 2020 Nov 26.

Abstract

Background: Air bubble intrusion through transseptal sheaths during left atrial (LA) catheter ablation can cause cerebral embolisms, especially when using complex-shape catheters. This study aimed to compare the incidence of silent cerebral events (SCEs) after atrial fibrillation (AF) catheter ablation using a mini-basket catheter (IntellaMap Orion; Boston Scientific) between the following groups: group SP, strict prevention of LA air intrusion and group CP, conventional air intrusion prevention.

Methods: We enrolled 123 consecutive AF patients (group SP, n = 61 and group CP, n = 62) who underwent brain magnetic resonance imaging after a local-impedance-guided ablation using one mini-basket catheter and one circular mapping catheter. The preventive strategy in group SP included (a) the insertion of the mini-basket catheter into the transseptal sheaths in a container filled with heparinized saline and (b) no exchange of all catheters over the sheaths.

Results: SCEs were detected in 67 patients (54.5%), and the incidence of SCEs did not significantly differ between groups SP and CP (55.7% vs 53.2%; P = .780). A multivariate analysis demonstrated that an older age, non-paroxysmal AF, and radiofrequency (RF) power output were independent positive predictors of SCEs (odds ratios: 1.079, 5.613, and 1.405; P = .005, <.001, and .012). On the follow-up MR imaging, 83.5% of the SCEs in group SP and 87.7% in group CP disappeared (P = .398).

Conclusions: Strict prevention of LA air intrusion may have no additional effect for reducing the incidence of SCEs after local impedance-guided AF ablation using a mini-basket catheter. An older age, non-paroxysmal AF, and high-power RF applications may increase the risk of SCEs.

Keywords: air intrusion; atrial fibrillation; catheter ablation; mini-basket catheter; silent cerebral event.

MeSH terms

  • Aged
  • Cardiac Catheterization / instrumentation*
  • Catheter Ablation / instrumentation
  • Catheter Ablation / methods*
  • Embolism, Air / prevention & control*
  • Equipment Design
  • Female
  • Humans
  • Magnetic Resonance Imaging
  • Male
  • Retrospective Studies
  • Stroke / prevention & control*