Compatibility of endoclips in the gastrointestinal tract with magnetic resonance imaging

Sci Rep. 2020 Oct 6;10(1):16537. doi: 10.1038/s41598-020-73726-5.

Abstract

There are no clear guidelines on the compatibility between endoclips that remain in the gastrointestinal (GI) tract and magnetic resonance imaging (MRI). The purpose of this study was to investigate the effect of 3T (T) MRI on endoclips placed in excised pig tissues. Two types of endoclips were assessed: Olympus EZ (HX-610-135L) and QuickClip Pro (HZ-202LR). We assessed tissue damage or perforation and detachment of endoclips under 3T MRI magnetic field. We also evaluated the magnitude of force required to detach the endoclips from the porcine tissue. We measured the magnetic force acting on the Olympus EZ clips. QuickClip Pro clips were used as a control in this study. There was no tissue damage and no detachment of the endoclips (Olympus EZ and QuickClip Pro) during 3T MRI. The force required to detach the Olympus EZ clips ranged from 0.9 to 3.0 N. The translational magnetic force acting on the endoclips was 3.18 × 10-3 N. Ex vivo experiments showed that the magnetic field generated by 3 MRI did not cause tissue damage or perforation and did not detach the endoclips. Olympus EZ clips and QuickClip Pro clips in the GI tract appear to be safe during 3T MRI.

MeSH terms

  • Animals
  • Equipment Design
  • Gastrointestinal Tract / diagnostic imaging*
  • Magnetic Fields / adverse effects*
  • Magnetic Resonance Imaging / adverse effects*
  • Magnetic Resonance Imaging / methods
  • Surgical Instruments / adverse effects*
  • Swine