Commissioning of total body irradiation using plastic bead bags

J Radiat Res. 2020 Nov 16;61(6):959-968. doi: 10.1093/jrr/rraa072.

Abstract

The goal of total body irradiation (TBI) is to deliver a dose to the whole body with uniformity within ±10%. The purpose of this study was to establish the technique of TBI using plastic bead bags. A lifting TBI bed, Model ORP-TBI-MN, was used. The space between the patient's body and the acrylic walls of the bed was filled with polyacetal bead bags. Patients were irradiated by a 10 MV photon beam with a source to mid-plane distance of 400 cm. The monitor unit (MU) was calculated by dose-per-MU, tissue-phantom-ratio and a spoiler factor measured in solid water using an ionization chamber. The phantom-scatter correction factor, off-center ratio and the effective density of the beads were also measured. Diode detectors were used for in vivo dosimetry (IVD). The effective density of the beads was 0.90 ± 0.09. The point doses calculated in an I'mRT phantom with and without heterogeneity material showed good agreement, with measurements within 3%. An end-to-end test was performed using a RANDO phantom. The mean ± SD (range) of the differences between the calculated and IVD-measured mid-plane doses was 1.1 ± 4.8% (-5.9 to 5.0%). The differences between the IVD-measured doses and the doses calculated with Acuros XB of the Eclipse treatment planning system (TPS) were within 5%. For two patients treated with this method, the differences between the calculated and IVD-measured doses were within ±6% when excluding the chest region. We have established the technique of TBI using plastic bead bags. The TPS may be useful to roughly estimate patient dose.

Keywords: in vivo dosimetry; Acuros XB; total body irradiation; treatment planning.

MeSH terms

  • Algorithms
  • Equipment Design
  • Humans
  • Particle Accelerators
  • Phantoms, Imaging*
  • Photons / therapeutic use
  • Plastics
  • Radiation Dosage
  • Radiometry
  • Radiotherapy Dosage*
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Reproducibility of Results
  • Tomography, X-Ray Computed
  • Water
  • Whole-Body Irradiation / methods*

Substances

  • Plastics
  • Water