Study of Tidal Volume and Positive End-Expiratory Pressure on Alveolar Recruitment Using Spiro Dynamics in Mechanically Ventilated Patients

Anesth Essays Res. 2020 Jan-Mar;14(1):154-159. doi: 10.4103/aer.AER_10_20. Epub 2020 Jun 22.

Abstract

Background and aims: Ventilator setting in the intensive care unit patients is a topic of debate and setting of tidal volume (TV) should be patient-specific based on lung mechanics. In this study, we have evaluated to develop optimal ventilator strategies through continuous and thorough monitoring of respiratory mechanics during ongoing ventilator support to prevent alveolar collapse and alveolar injury in mechanically ventilated patients.

Methods: In our monocentric, randomized, observational study, we had recruited 60 patients and divided them into two groups of 30 each. In Group 1 patients, TV and positive end-expiratory pressure (PEEP) were set according to pressure-volume (P/V) curve obtained by the mechanical ventilator in a conventional manner (control group), and in Group 2, TV and PEEP were set according to P/V curve obtained by the mechanical ventilator using intratracheal catheter. PEEP and TV were set accordingly. TV, PEEP, and PaO2/FiO2 (P/F) ratio at days 1, 3, and 7, mortality within 7 days and mortality within 28 days were measured in each group and compared.

Results: We found a significant difference between PEEP and P/F ratio in both groups while intragroup comparison at days 1, 3, and 7. After the intergroup comparison of Group 1 and 2, we observed a significant difference of PEEP and P/F ratio between the groups at day 7 and not on day 1 or 3.

Conclusion: This study concludes that optimal PEEP is more accurate using an intratracheal catheter than the conventional method of deciding ventilator setting. Hence, it is recommended to use intratracheal catheter to obtain more accurate ventilator settings.

Keywords: Intratracheal catheter; mechanical ventilation; positive end-expiratory pressure; tidal volume.