Preparation and characterisation of bifunctional surface-modified silicone catheter in lumen

J Glob Antimicrob Resist. 2020 Dec:23:46-54. doi: 10.1016/j.jgar.2020.07.019. Epub 2020 Aug 11.

Abstract

Objectives: The purpose of this study was to evaluate the coating of antimicrobial peptides (AMPs) and polyvinylpyrrolidone (PVP) to the surface of a silicone catheter to reduce bacterial growth and to increase hydrophilicity, respectively.

Methods: Surface characterisation was performed on bare silicone, AMP-coated, PVP-coated and AMP + PVP-coated silicone catheters using attenuated total reflectance-infrared (ATR-IR) spectroscopy, X-ray photoelectron spectroscopy (XPS) and water contact angle. Antibacterial activity, antibacterial biofilm growth and XTT assay were performed on bare silicone, AMP-coated, PVP-coated and AMP + PVP-coated silicone catheters. Statistical analysis was performed by one-way ANOVA.

Results: The water contact angle of the AMP + PVP-coated silicone catheter was 21.37 ± 2.17° compared with 107.23 ± 0.96°, 74.40 ± 1.76° and 20.77 ± 0.32° for bare silicone, AMP-coated and PVP-coated silicone catheters. Based on in vitro antimicrobial tests, the AMP + PVP-coated silicone catheter had 6.2, 2.2 and 2.5 greater antibacterial activity than that of the bare silicone catheter against Escherichia coli, Staphylococcus aureus and Pseudomonas aeruginosa, respectively. Moreover, bacterial biofilm growth on the surface of the AMP + PVP-coated silicone catheter was minimal as characterised by scanning electron microscopy. MTT assay showed that bare silicone, AMP-coated, AMP + PVP-coated and PVP-coated silicone catheters were non-cytotoxic to 3T3 and human colon cancer (Caco-2) cells.

Conclusions: This work demonstrates that AMP + PVP-coated silicone catheters have potential clinical application prospects with improved hydrophilicity, excellent biocompatibility, antibacterial activity and a certain antibacterial biofilm effect.

Keywords: Antibacterial; Antibiofilm; Antimicrobial peptide; Lubricated; Polyvinylpyrrolidone; Silicone catheter.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biofilms
  • Caco-2 Cells
  • Catheters
  • Humans
  • Silicones*
  • Staphylococcus aureus*

Substances

  • Silicones