Transcatheter InterAtrial Shunt Device for the treatment of heart failure: Rationale and design of the pivotal randomized trial to REDUCE Elevated Left Atrial Pressure in Patients with Heart Failure II (REDUCE LAP-HF II)

Am Heart J. 2020 Aug:226:222-231. doi: 10.1016/j.ahj.2019.10.015. Epub 2019 Nov 11.

Abstract

Background: A randomized, sham-controlled trial in patients with heart failure (HF) and left ventricular ejection fraction (LVEF) ≥40% demonstrated reductions in pulmonary capillary wedge pressure (PCWP) with a novel transcatheter InterAtrial Shunt Device (IASD). Whether this hemodynamic effect will translate to an improvement in cardiovascular outcomes and symptoms requires additional study.

Study design and objectives: REDUCE Elevated Left Atrial Pressure in Patients with Heart Failure II (REDUCE LAP HF-II) is a multicenter, prospective, randomized, sham-controlled, blinded trial designed to evaluate the clinical efficacy of the IASD in symptomatic HF and elevated left atrial pressures. Up to 608 HF patients age ≥ 40 years with LVEF ≥40%, PCWP ≥25 mm Hg during supine ergometer exercise, and PCWP ≥5 mm Hg higher than right atrial pressure will be randomized 1:1 to the IASD versus sham control. Key exclusion criteria include hemodynamically significant valvular disease, evidence of pulmonary arterial hypertension, and right heart dysfunction. The primary endpoint is a hierarchical composite, analyzed by the Finkelstein-Schoenfeld methodology, that includes (1) cardiovascular mortality or first nonfatal ischemic stroke through 12 months; (2) total (first plus recurrent) HF hospitalizations or healthcare facility visits for intravenous diuretics up to 24 months, analyzed when the last randomized patient completes 12 months of follow-up; and (3) change in Kansas City Cardiomyopathy Questionnaire overall summary score from baseline to 12 months. Follow-up echocardiography will be performed at 6, 12, and 24 months to evaluate shunt flow and cardiac chamber size/function. Patients will be followed for a total of 5 years after the index procedure.

Conclusions: REDUCE LAP-HF II is designed to evaluate the clinical efficacy of the IASD device in patients with symptomatic HF with elevated left atrial pressure and LVEF ≥40%.

Publication types

  • Clinical Trial Protocol
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cardiac Catheterization
  • Double-Blind Method
  • Heart Failure / physiopathology
  • Heart Failure / surgery*
  • Humans
  • Multicenter Studies as Topic
  • Prospective Studies
  • Prostheses and Implants*
  • Prosthesis Implantation / methods*
  • Randomized Controlled Trials as Topic / methods*
  • Ventricular Function, Left