Comparison of optimal screw configurations in two locking plate systems for proximal humerus fixation - a finite element analysis study

Clin Biomech (Bristol, Avon). 2020 Aug:78:105097. doi: 10.1016/j.clinbiomech.2020.105097. Epub 2020 Jun 27.

Abstract

Background: Management of proximal humerus fractures is challenging, especially in elderly. Locking plating is a common surgical treatment option. The Proximal Humerus Internal Locking System (plate-A) has shown to lower complication rates compared to conventional plates, but is associated with impingement risk, which could be avoided using Peri-articular Proximal Humerus Plate (plate-B). Nevertheless, biomechanical performance and optimal screw configuration of plate-B is unknown. The aim of this study was to evaluate different screw configurations of plate-B and compare with plate-A using finite element analyses.

Methods: Twenty-six proximal humerus models were osteotomised to create unstable three-part fractures, fixed with either of the two plates, and tested under three anatomical loading conditions using a previous established and validated finite element simulation framework. Various clinically relevant screw configurations were investigated for both plates and compared based on the predicted peri-implant bone strain, being a validated surrogate of cyclic cut-out failure.

Findings: Besides increasing the number of screws, the placement of the posterior screws in combination with the calcar screw in the plate-B significantly decreased the predicted failure risk. Generally, plate-A had a lower predicted failure risk than plate-B.

Interpretation: The posterior and calcar screws may be prioritized in plate-B. Compared to plate-A, the more distal positioning, less purchase in the posterior aspect and a smaller screw spread due to not fitting of the most distal calcar screw in most investigated subjects led to a significantly higher predicted failure risk for most plate-B configurations. The findings of the simulations study require clinical corroboration.

Keywords: Finite element analysis; Fracture fixation; Locking plates; Proximal humerus; Shoulder.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Biomechanical Phenomena
  • Bone Plates
  • Bone Screws*
  • Computer Simulation
  • Female
  • Finite Element Analysis*
  • Fracture Fixation, Internal / instrumentation*
  • Humans
  • Male
  • Shoulder Fractures / surgery*