Conical ring array detector for large depth of field photoacoustic macroscopy

Biomed Opt Express. 2020 Apr 9;11(5):2461-2475. doi: 10.1364/BOE.386585. eCollection 2020 May 1.

Abstract

Photoacoustic microscopy and macroscopy (PAM) using focused detector scanning are emerging imaging methods for biological tissue, providing high resolution and high sensitivity for structures with optical absorption contrast. However, achieving a constant lateral resolution over a large depth of field for deeply penetrating photoacoustic macroscopy is still a challenge. In this work, a detector design for scanning photoacoustic macroscopy is presented. Based on simulation results, a sensor array geometry is developed and fabricated that consists of concentric ring elements made of polyvinylidene fluoride (PVDF) film in a geometry that combines a centered planar ring with several inclined outer ring elements. The reconstruction algorithm, which uses dynamic focusing and coherence weighting, is explained and its capability to reduce artefacts occurring for single element conical sensors is demonstrated. Several phantoms are manufactured to evaluate the performance of the array in experimental measurements. The sensor array provides a constant axial and lateral resolution of 95 µm and 285 µm, respectively, over a depth of field of 20 mm. The depth of field corresponds approximately to the maximum imaging depth in biological tissue, estimated from the sensitivity of the array. With its ability to achieve the maximum resolution even with a very small scanning range, the array is believed to have applications in the imaging of limited regions of interest buried in biological tissue.