A Three-Way Accuracy Comparison of the Dexcom G5, Abbott Freestyle Libre Pro, and Senseonics Eversense Continuous Glucose Monitoring Devices in a Home-Use Study of Subjects with Type 1 Diabetes

Diabetes Technol Ther. 2020 Nov;22(11):846-852. doi: 10.1089/dia.2019.0449.

Abstract

Background: There is a dearth of comparative accuracy studies of continuous glucose monitoring (CGM) devices in the home-use setting, and none with the Eversense implantable CGM. Methods: We evaluated the accuracy of the Dexcom G5, Abbott Freestyle Libre Pro, and Senseonics Eversense during a 6-week free-living home-use bionic pancreas study involving 23 subjects with type 1 diabetes who wore all three devices concurrently. The primary outcome was the mean absolute relative difference (MARD) between CGM readings and point-of-care (POC) plasma-glucose (PG) values obtained approximately twice daily by the subjects. We compared PG values with CGM readings when available from all three CGMs in the 5 min preceding the PG values (n = 829 sets). Since the Libre Pro records readings every 15 min, we also did a two-way comparison between the G5 and the Eversense with a higher number of comparisons (n = 2277 sets). Results: All three CGM systems produced higher average MARDs than during in-clinic studies. However, since all three CGM systems were worn by the same individuals and used the same meter for comparator PG measurements, we could directly compare their performances. In the three-way comparison, Eversense achieved the lowest nominal MARD (14.8%) followed by Dexcom G5 (16.3%) and Libre Pro (18.0%) (Eversense vs. Libre Pro P = 0.004, other comparisons P = NS). There was a statistically significant difference (P = 0.008) in the two-way comparison of the MARDs for Eversense (15.1%) and G5 (16.9%). Conclusions: The point accuracy of the Eversense was better than two other CGMs when compared with POC PG values.

Keywords: Abbott Libre; Accuracy; Continuous glucose monitoring; Dexcom G5; Senseonics Eversense.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Glucose / analysis
  • Blood Glucose Self-Monitoring / instrumentation*
  • Diabetes Mellitus, Type 1* / diagnosis
  • Humans
  • Pancreas, Artificial
  • Reproducibility of Results

Substances

  • Blood Glucose