4D Volume Intracardiac Echocardiography for Intraprocedural Guidance of Transcatheter Left Atrial Appendage Closure

J Atr Fibrillation. 2019 Dec 31;12(4):2200. doi: 10.4022/jafib.2200. eCollection 2019 Dec.

Abstract

Background: Fluoroscopy and transesophageal echocardiography (TEE) are used to guide transcatheter left atrial appendage (LAA) closure in patients with atrial fibrillation to prevent thromboembolic events. This study examines whether real-time three-dimensional volume ICE guidance (4D volume ICE) can be used as an alternative to TEE during LAA closure (LAAC).

Methods and results: Fifteen patients with atrial fibrillation (AF), who had high risk for stroke and contraindication for long-term warfarin therapy, were enrolled in the study. The WATCHMAN device was used for transcatheter LAAC under fluoroscopy. LAA and device sizing was performed using TEE and volume ICE guidance from the right heart. Intraprocedural ICE measurements were consistent with TEE; LAA maximal width and depth, and maximal diameter of the implanted device were moderately correlated (Pearson's coefficient: 0.63, 0.65, and 0.71 respectively; p<0.05) with good agreement (bias: -0.03 cm, -0.07 cm, and 0.003 cm respectively). The average imaging success rate, defined by the number of patients with all the required intraprocedural measurements, was 67% for ICE and 100% for TEE. The WATCHMAN device was successfully implanted in all patients with a device to patient ratio of 1.33.

Conclusions: 4D volume ICE can be used as an intraprocedural sizing and guidance tool for transcatheter LAAC with measurements comparable to TEE. Challenging patient characteristics significantly degrade the diagnostic image quality when imaging from the right heart. Standardized workflow with proper patient selection and optimal preprocedural planning may improve the diagnostic quality of volume ICE guidance for transcatheter LAAC procedure.

Keywords: ICE; LAA Closure; TEE; WATCHMAN.