Comparing LigasureTM Exact dissector with other energy devices in total thyroidectomy: a pilot study

Gland Surg. 2020 Apr;9(2):271-277. doi: 10.21037/gs.2020.02.05.

Abstract

Background: Hemostasis is fundamental in thyroid surgery. The aim of this randomized active comparator-controlled study was to compare the effectiveness of LigasureTM Exact to previously existing sealing devices in patients undergoing thyroidectomy.

Methods: A total of 180 patients were randomly allocated into 4 groups according to the hemostatic device used: Focus Harmonic Scalpel® (FHS), LigasureTM Small Jaw (LSJ), Thunderbeat® (TB), and LigasureTM Exact (LE). We recorded epidemiologic and biochemical data, operative difficulty, postoperative vocal alteration and pain, complications, operating time, grams of gland excised per minute, and hospitalization.

Results: Differences (P<0.001) were observed concerning duration of surgery, operative difficulty and quantity of gland removed per minute. Concerning duration of surgery it is shorter for LE followed by TB and then by both FHS and LSJ. Concerning the quantity of thyroid removed per minute LE leads, followed by FHS and TB. The least amount of gland removed per minute is by LSJ. Finally, operative difficulty was less for LE, while it was equal for all other three devices.

Conclusions: Energy devices are now frequently used in thyroidectomy. Comparing these devices seems to qualify them as equal concerning post-operative complications, hemostasis, and acute inflammatory reaction. However, LigasureTM Exact Dissector seems to have better results concerning duration of surgery, quantity of thyroid tissue removed per minute and intraoperative difficulty.

Keywords: Focus Harmonic Scalpel; Ligasure Exact; Ligasure Small Jaw; Thyroid surgery; thyroidectomy.