Real-World Usage of the WavelinQ EndoAVF System

Ann Vasc Surg. 2021 Jan:70:116-122. doi: 10.1016/j.avsg.2020.05.006. Epub 2020 May 15.

Abstract

Background: Dependent on existing deep to superficial perforating venous branches, the WavelinQ EndoAVF System is a novel technique used to create an arteriovenous fistula (AVF) between ulnar or radial veins and concomitant arteries for dialysis access. We sought to examine a single center's success rates and short-term follow-up using this device.

Methods: All consecutive patients undergoing placement of a WavelinQ AVF from October 2018 to July 2019 were included. Preoperative/intraoperative variables including demographics, preoperative/postoperative duplex ultrasonography, success rate of procedure, and subsequent endovascular/surgical procedures were obtained. Descriptive statistics and comparison of groups requiring subsequent intervention were performed.

Results: Thirty-five patients underwent placement of the WavelinQ AVF, with 32 (91%) patients having at least one documented follow-up. These patients were predominantly male (23/32, 72%) with an average age of 60.2 and 23 of 32 (72%) patients were on dialysis. Initial fistula creation success rate was 100%. Average procedural length was 120 min, fluoroscopy time 9.6 min, and contrast usage 52.2 mL. Eight of 32 (25%) patients had perioperative complications (3 hematomas, 3 contrast extravasations, 1 resolved vessel spasm all resolving spontaneously, and 1 pseudoaneurysm requiring surgical repair). Thirteen of 32 (41%) patients underwent subsequent endovascular interventions to assist with maturation [9/32 (28%) branch coiling, 5/32 (16%) angioplasty/stenting, and 3/32 (9%) access thrombectomy] and 4 of 32 (13%) patients required subsequent surgical interventions (1 pseudoaneurysm repair, 1 revision of fistula, and 2 definitive AVF creation in thrombosed grafts). The majority of accesses (30/32, 94%) were ulnar-ulnar fistulas and overall patency at average follow-up of 73 days was 88% (28/32) with average brachial artery inflow volume of 1,078 cc/min and average cephalic vein (18/32) outflow volume of 447 cc/min. Eleven of 23 (48%) patients on dialysis were successfully using the EndoAVF at follow-up.

Conclusions: The WavelinQ AVF system has a high initial procedural success rate, although a significant portion of patients require subsequent endovascular procedures to aid in maturation. Further work on determining factors predictive of need for reintervention is necessary.

MeSH terms

  • Arteriovenous Shunt, Surgical / adverse effects
  • Arteriovenous Shunt, Surgical / instrumentation*
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis*
  • Female
  • Graft Occlusion, Vascular / etiology
  • Graft Occlusion, Vascular / physiopathology
  • Graft Occlusion, Vascular / therapy
  • Humans
  • Male
  • Middle Aged
  • Radial Artery / diagnostic imaging
  • Radial Artery / physiopathology
  • Radial Artery / surgery*
  • Renal Dialysis*
  • Retrospective Studies
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • Ulnar Artery / diagnostic imaging
  • Ulnar Artery / physiopathology
  • Ulnar Artery / surgery*
  • Upper Extremity / blood supply*
  • Vascular Patency
  • Veins / diagnostic imaging
  • Veins / physiopathology
  • Veins / surgery*