MR-based PET attenuation correction using a combined ultrashort echo time/multi-echo Dixon acquisition

Med Phys. 2020 Jul;47(7):3064-3077. doi: 10.1002/mp.14180. Epub 2020 May 11.

Abstract

Purpose: To develop a magnetic resonance (MR)-based method for estimation of continuous linear attenuation coefficients (LACs) in positron emission tomography (PET) using a physical compartmental model and ultrashort echo time (UTE)/multi-echo Dixon (mUTE) acquisitions.

Methods: We propose a three-dimensional (3D) mUTE sequence to acquire signals from water, fat, and short T2 components (e.g., bones) simultaneously in a single acquisition. The proposed mUTE sequence integrates 3D UTE with multi-echo Dixon acquisitions and uses sparse radial trajectories to accelerate imaging speed. Errors in the radial k-space trajectories are measured using a special k-space trajectory mapping sequence and corrected for image reconstruction. A physical compartmental model is used to fit the measured multi-echo MR signals to obtain fractions of water, fat, and bone components for each voxel, which are then used to estimate the continuous LAC map for PET attenuation correction.

Results: The performance of the proposed method was evaluated via phantom and in vivo human studies, using LACs from computed tomography (CT) as reference. Compared to Dixon- and atlas-based MRAC methods, the proposed method yielded PET images with higher correlation and similarity in relation to the reference. The relative absolute errors of PET activity values reconstructed by the proposed method were below 5% in all of the four lobes (frontal, temporal, parietal, and occipital), cerebellum, whole white matter, and gray matter regions across all subjects (n = 6).

Conclusions: The proposed mUTE method can generate subject-specific, continuous LAC map for PET attenuation correction in PET/MR.

Keywords: MR-based attenuation correction; MRAC; PET attenuation correction; PET/MR; continuous LAC; multi-echo Dixon; ultrashort echo time (UTE).

MeSH terms

  • Humans
  • Image Processing, Computer-Assisted*
  • Magnetic Resonance Imaging
  • Phantoms, Imaging
  • Positron-Emission Tomography*
  • Tomography, X-Ray Computed