In order to better understand bone conduction sound propagation across the skull, three-dimensional (3D) wave propagation on the skull surface was studied, along with its dependence on stimulation direction and location of a bone conduction hearing aid (BCHA) actuator. Experiments were conducted on five Thiel embalmed whole head cadaver specimens. Stimulation, in the 0.1-10 kHz range, was sequentially applied at the forehead and mastoid via electromagnetic actuators from commercial BCHAs, supported by a 5-N steel band. The head response was quantified by sequentially measuring the 3D motion of ∼200 points (∼15-20 mm pitch) across the ipsilateral, top, and contralateral skull surface via a 3D laser Doppler vibrometer (LDV) system, guided by a robotic positioner. Low-frequency stimulation (<1 kHz) resulted in a spatially complex rigid-body-like motion of the skull that depended on both the stimulation condition and head support. The predominant motion direction was only 5-10 dB higher than other components below 1 kHz, with no predominance at higher frequencies. Sound propagation direction across the parietal plates did not coincide with stimulation location, potentially due to the head base and forehead remaining rigid-like at higher frequencies and acting as a large source for the deformation patterns across the parietal sections.