Aims: To comparatively analyse outcomes after ARGUS classic and ArgusT adjustable male sling implantation in a real-world setting.

Methods: Inclusion criteria encompassed: non-neurogenic, moderate-to-severe stress urinary incontinence (≥2 pads), implantation of an ARGUS classic or ArgusT male sling between 2010 and 2012 in a high-volume center (>150 previous implantations). Functional outcomes were assessed using daily pad usage, 24-hour pad testing, and International Consultation on Incontinence (ICIQ-SF) questionnaires. Continence was defined as zero pads per 24 hours. Complications were graded using Clavien-Dindo scale. For multivariate analysis, binary logistic regression models were used (P < .05).

Results: A total of 106 patients (n = 74 [ARGUS classic], n = 32 [ArgusT]) from four centers were eligible. Median follow-up was 44.0 months (24-64). Patient cohorts were well-balanced. We observed a statistical trend in favor of ARGUS classic regarding postoperative urine loss based on standardized 24-hour pad test (71 ± 162 g [ARGUS classic] vs 160 ± 180 g [ArgusT]); P = .066]. Dry rates were 33.3% (ARGUS classic) and 11.8% (ArgusT, P = .114). In multivariable analysis, ArgusT male sling could not be confirmed as an independent predictor of decreased success (OR, 0.587; 95% CI; 0.166-2.076, P = .408). There were no Clavien IV and V complications. Inguinal pain was significantly higher after ArgusT implantation (P = .033). Explantation rates were higher for ArgusT compared to ARGUS classic (14.0 vs 23.3%; P = .371). Longer device survival for the ARGUS classic male sling was observed in Kaplan-Meier analysis (P = .198).

Conclusions: In the largest comparative analysis of ARGUS classic and ArgusT male sling with the longest follow-up to date, we observed low continence rates with superior functional outcomes and decreased explantation rates after ARGUS classic implantation.

Keywords: adjustable male sling; male urinary incontinence; patient-reported outcomes; post-prostatectomy incontinence.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Device Removal / statistics & numerical data*
  • Humans
  • Male
  • Middle Aged
  • Postoperative Complications / epidemiology
  • Postoperative Complications / surgery*
  • Prostatectomy*
  • Suburethral Slings*
  • Surveys and Questionnaires
  • Treatment Outcome
  • Urinary Incontinence, Stress / surgery*
  • Urologic Surgical Procedures, Male / instrumentation
  • Urologic Surgical Procedures, Male / methods*