Dosimetric evaluation of the Leksell GammaPlan Convolution dose calculation algorithm

Phys Med Biol. 2020 Feb 12;65(4):045011. doi: 10.1088/1361-6560/ab64b7.

Abstract

The dosimetric accuracy of the Leksell GammaPlan Convolution calculation algorithm was evaluated through comparison with corresponding Monte Carlo (MC) dosimetric results. MC simulations were based on generated sector phase space files for the 4 mm, 8 mm and 16 mm collimator sizes, using a previous comprehensive Gamma Knife Perfexion source model and validated using film dosimetry. Test cases were designed for the evaluation of the Convolution algorithm involving irradiation of homogeneous and inhomogeneous phantom geometries mimicking clinical cases, with radiation fields created using one sector (single sector), all sectors with the same (single shot) or different (composite shot) collimator sizes. Dose calculations using the Convolution algorithm were found to be in excellent agreement (gamma pass rate greater than 98%, applying 1%/1 mm local dose difference and distance agreement criteria), with corresponding MC calculations, indicating the accuracy of the Convolution algorithm in homogeneous and heterogeneous model geometries. While of minor clinical importance, large deviations were observed for the voxels laying inside air media. The calculated beam on times using the Convolution algorithm were found to increase (up to 7%) relative to the TMR 10 algorithm currently used in clinical practice, especially in a test case mimicking a brain metastasis close to the skull, in excellent agreement with corresponding MC calculations.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Brain Neoplasms / diagnostic imaging
  • Brain Neoplasms / secondary
  • Humans
  • Monte Carlo Method
  • Phantoms, Imaging
  • Radiation Dosage*
  • Radiometry
  • Radiosurgery*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted