A Randomized, Placebo-Controlled Double-Blind Trial of a Closed-Loop Glucagon System for Postbariatric Hypoglycemia

J Clin Endocrinol Metab. 2020 Apr 1;105(4):e1260-e1271. doi: 10.1210/clinem/dgz197.

Abstract

Background: Postbariatric hypoglycemia (PBH) can threaten safety and reduce quality of life. Current therapies are incompletely effective.

Methods: Patients with PBH were enrolled in a double-blind, placebo-controlled, crossover trial to evaluate a closed-loop glucose-responsive automated glucagon delivery system designed to reduce severe hypoglycemia. A hypoglycemia detection and mitigation algorithm was embedded in the artificial pancreas system connected to a continuous glucose monitor (CGM, Dexcom) driving a patch infusion pump (Insulet) filled with liquid investigational glucagon (Xeris) or placebo (vehicle). Sensor/plasma glucose responses to mixed meal were assessed during 2 study visits. The system delivered up to 2 doses of study drug (300/150 μg glucagon or equal-volume vehicle) if triggered by the algorithm. Rescue dextrose was given for plasma glucose <55 mg/dL or neuroglycopenia.

Results: Twelve participants (11 females/1 male, age 52 ± 2, 8 ± 1 years postsurgery, mean ± SEM) completed all visits. Predictive hypoglycemia alerts prompted automated drug delivery postmeal, when sensor glucose was 114 ± 7 vs 121 ± 5 mg/dL (P = .39). Seven participants required rescue glucose after vehicle but not glucagon (P = .008). Five participants had severe hypoglycemia (<55 mg/dL) after vehicle but not glucagon (P = .03). Nadir plasma glucose was higher with glucagon vs vehicle (67 ± 3 vs 59 ± 2 mg/dL, P = .004). Plasma glucagon rose after glucagon delivery (1231 ± 187 vs 16 ± 1 pg/mL at 30 minutes, P = .001). No rebound hyperglycemia occurred. Transient infusion site discomfort was reported with both glucagon (n = 11/12) and vehicle (n = 10/12). No other adverse events were observed.

Conclusion: A CGM-guided closed-loop rescue system can detect imminent hypoglycemia and deliver glucagon, reducing severe hypoglycemia in PBH.

Clinical trials registration: NCT03255629.

Keywords: closed-loop; gastric bypass; glucagon; hypoglycemia.

Publication types

  • Randomized Controlled Trial
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Bariatric Surgery / adverse effects*
  • Cross-Over Studies
  • Double-Blind Method
  • Female
  • Follow-Up Studies
  • Gastrointestinal Agents / administration & dosage*
  • Glucagon / administration & dosage*
  • Humans
  • Hypoglycemia / drug therapy*
  • Hypoglycemia / etiology
  • Hypoglycemia / pathology
  • Male
  • Middle Aged
  • Obesity, Morbid / surgery*
  • Prognosis

Substances

  • Gastrointestinal Agents
  • Glucagon

Associated data

  • ClinicalTrials.gov/NCT03255629