Antibody immobilization on swollen polystyrene tubes for the radioimmunological determination of total human serum IgE

Clin Chim Acta. 1988 Aug 15;176(1):91-9. doi: 10.1016/0009-8981(88)90178-7.

Abstract

A new technique for preparing activated polystyrene tubes was developed involving the controlled swelling of plastic. The quantity of antibodies immobilized on these tubes, and consequently the quantity of bound IgE, was considerably increased. Calibration curves were plotted and the determination of total IgE in 200 serum samples confirmed the expected increase in sensitivity compared to a standard method. The manufacturing procedure was simple and could easily be automated. Thus, swollen polystyrene tubes constitute an advantageous solid stage in the assay of total IgE, which can improve the sensitivity of various radioimmunological determinations.

MeSH terms

  • Antibodies
  • Humans
  • Immunoglobulin E / analysis*
  • Polystyrenes
  • Radioimmunoassay / instrumentation
  • Radioimmunoassay / methods

Substances

  • Antibodies
  • Polystyrenes
  • Immunoglobulin E