Artificial Urinary Sphincter Cuff Size Predicts Outcome in Male Patients Treated for Stress Incontinence: Results of a Large Central European Multicenter Cohort Study

Int Neurourol J. 2019 Sep;23(3):219-225. doi: 10.5213/inj.1938032.016. Epub 2019 Sep 30.

Abstract

Purpose: The aim was to study the correlation between cuff size and outcome after implantation of an AMS 800 artificial urinary sphincter.

Methods: A total of 473 male patients with an AMS 800 sphincter implanted between 2012 and 2014 were analyzed in a retrospective multicenter cohort study performed as part of the Central European Debates on Male Incontinence (DOMINO) Project.

Results: Single cuffs were implanted in 54.5% and double cuffs in 45.5% of the patients. The cuffs used had a median circumference of 4.5 cm. Within a median follow of 18 months, urethral erosion occurred in 12.8% of the cases and was associated significantly more often with small cuff sizes (P<0.001). Multivariate analysis showed that, apart from cuff size (P=0.03), prior irradiation (P<0.001) and the penoscrotal approach (P=0.036) were associated with an increased erosion rate. Continence rate tended to be highest with median cuff sizes (4-5.5 cm).

Conclusion: Apart from irradiation and the penoscrotal approach, small cuff size is a risk factor for urethral erosion. Results are best with cuff sizes of 4.5-5.5 cm.

Keywords: AMS 800; Artificial urinary sphincter; Cuff size; Male; Single or double cuff; Stress urinary incontinence.