High Incidence of Intraoperative Fractures With a Specific Cemented Stem Following Intracapsular Displaced Hip Fracture

J Arthroplasty. 2020 Feb;35(2):485-489. doi: 10.1016/j.arth.2019.09.017. Epub 2019 Sep 17.

Abstract

Background: To reduce costs of orthopedic implants, the government decided to standardize implants used across different specialties in a group of hospitals located in the same geographic area. The usual cemented stem used in the context of intracapsular displaced geriatric hip fractures was replaced by another stem. Abnormal intraoperative calcar and trochanteric fractures were noted. The purpose of this study is to determine the incidence of intraoperative periprosthetic fractures following an intracapsular displaced hip fracture treated with this specific cemented stem compared to the previous implant.

Methods: This is a retrospective cohort study comparing an historic cohort of hip fractures treated with the OmniFit EON (Stryker, Kalamazoo, MI) cemented stem with a new cohort of patients who received the Corail (DePuy Synthes, Warsaw, IN) cemented stem. Four orthopedic surgeons reviewed operative reports and postoperative radiographs.

Results: The treatment group included 348 patients who received the Corail stem. The control group included 77 patients. The 2 groups had similar baseline characteristics (P > .05) except for the presence of dementia. Incidence of intraoperative calcar or greater trochanteric fracture was 15.5% for the Corail group and 2.7% for the control group (P < .05). No patient-related factors or surgeon-related factors were related to a higher number of fractures in the treatment group (P > .05).

Conclusion: The Corail cemented stem presents an abnormal number of iatrogenic intraoperative fractures following displaced femoral neck fracture in our geriatric population. No external factor seems to explain this high number of fractures. Implant design should be questioned.

Level of evidence: III.

Keywords: calcar fracture; cemented stem; femoral neck fracture; hemiarthroplasty; hip fracture.

MeSH terms

  • Aged
  • Arthroplasty, Replacement, Hip* / adverse effects
  • Femoral Neck Fractures* / surgery
  • Hemiarthroplasty*
  • Hip Fractures* / epidemiology
  • Hip Fractures* / etiology
  • Hip Fractures* / surgery
  • Hip Prosthesis* / adverse effects
  • Humans
  • Incidence
  • Retrospective Studies
  • Treatment Outcome