Capillary column gas chromatographic method using electron-capture detection for the simultaneous determination of nicardipine and its pyridine metabolite II in plasma

J Chromatogr. 1987 Mar 20;415(1):65-73. doi: 10.1016/s0378-4347(00)83192-8.

Abstract

A rapid, specific and direct method based on capillary column gas chromatography with electron-capture detection is described for the simultaneous determination of nicardipine, a new calcium antagonist, and its pyridine metabolite II in human plasma. In this method, the nicardipine, its pyridine metabolite II and internal standard are extracted from the plasma and then partially purified by acid-base partitioning prior to the final injection onto the capillary column gas chromatograph for quantification by means of an electron-capture detector. The quantification limit of the method is 1 ng/ml of plasma for both nicardipine and its pyridine metabolite II. The coefficients of variation for nicardipine and the pyridine metabolite II at concentrations of 1-50 ng/ml are less than 7% and less than 9% (n = 4), respectively. The method has been validated against a previously developed high-performance liquid chromatographic method (sensitivity 5 ng/ml).

MeSH terms

  • Chromatography, Gas
  • Humans
  • Nicardipine / analogs & derivatives*
  • Nicardipine / blood*

Substances

  • 2-(N-benzyl-N-methylamino)ethyl methyl 2,6-dimethyl-4-(3-nitrophenyl)-3,5-pyridinedicarboxylate
  • Nicardipine