Electrical impedance tomography during spontaneous breathing trials and after extubation in critically ill patients at high risk for extubation failure: a multicenter observational study

Ann Intensive Care. 2019 Aug 13;9(1):88. doi: 10.1186/s13613-019-0565-0.

Abstract

Background: This study aims to assess the changes in lung aeration and ventilation during the first spontaneous breathing trial (SBT) and after extubation in a population of patients at risk of extubation failure.

Methods: We included 78 invasively ventilated patients eligible for their first SBT, conducted with low positive end-expiratory pressure (2 cm H2O) for 30 min. We acquired three 5-min electrical impedance tomography (EIT) records at baseline, soon after the beginning (SBT_0) and at the end (SBT_30) of SBT. In the case of SBT failure, ventilation was reinstituted; otherwise, the patient was extubated and two additional records were acquired soon after extubation (SB_0) and 30 min later (SB_30) during spontaneous breathing. Extubation failure was defined by the onset of post-extubation respiratory failure within 48 h after extubation. We computed the changes from baseline of end-expiratory lung impedance (∆EELI), tidal volume (∆Vt%), and the inhomogeneity index. Arterial blood was sampled for gas analysis. Data were compared between sub-groups stratified for SBT and extubation success/failure.

Results: Compared to SBT success (n = 61), SBT failure (n = 17) showed a greater reduction in ∆EELI at SBT_0 (p < 0.001) and SBT_30 (p = 0.001) and a higher inhomogeneity index at baseline (p = 0.002), SBT_0 (p = 0.003) and SBT_30 (p = 0.005). RR/Vt was not different between groups at baseline but was significantly greater at SBT_0 and SBT_30 in SBT failures, compared to SBT successes (p < 0.001 for both). No differences in ∆Vt% and arterial blood gases were observed between SBT success and failure. The ∆Vt%, ∆EELI, inhomogeneity index and arterial blood gases were not different between patients with extubation success (n = 39) and failure (n = 22) (p > 0.05 for all comparisons).

Conclusions: Compared to SBT success, SBT failure was characterized by more lung de-recruitment and inhomogeneity. Whether EIT may be useful to monitor SBT remains to be determined. No significant changes in lung ventilation, aeration or homogeneity related to extubation outcome occurred up to 30 min after extubation. Trial registration Retrospectively registered on clinicaltrials.gov (Identifier: NCT03894332; release date 27th March 2019).

Keywords: Electrical impedance tomography; Extubation failure; Mechanical ventilation; Post-extubation respiratory failure; Spontaneous breathing trial; Weaning.

Associated data

  • ClinicalTrials.gov/NCT03894332