Using EQ·PET to reduce reconstruction-dependent variations in [18F]FDG-PET brain imaging

Phys Med Biol. 2019 Aug 28;64(17):175002. doi: 10.1088/1361-6560/ab35b4.

Abstract

This study aims at assessing whether EANM harmonisation strategy combined with EQ·PET methodology could be successfully applied to harmonize brain 2-deoxy-2[18F]fluoro-D-glucose ([18F]FDG) positron emission tomography (PET) images. The NEMA NU 2 body phantom was prepared according to the EANM guidelines with an [18F]FDG solution. Raw PET phantom data were reconstructed with three different reconstruction protocols frequently used in clinical PET brain imaging: ([Formula: see text]) Ordered subset expectation maximization (OSEM) 3D with time of flight (TOF), 2 iterations and 21 subsets; ([Formula: see text]) OSEM 3D with TOF, 6 iterations and 21 subsets; and ([Formula: see text]) OSEM 3D with TOF, point spread function (PSF), and 8 iterations and 21 subsets. EQ·PET filters were computed as the Gaussian smoothing that best independently aligned the recovery coefficients (RCs) of reconstructions [Formula: see text] and [Formula: see text] with the RCs of the reference reconstruction, [Formula: see text]. The performance of the EQ·PET filter to reduce variations in quantification due to differences in reconstruction was investigated using clinical PET brain images of 35 early-onset Alzheimer's disease (EOAD) patients. Qualitative assessments and multiple quantitative metrics on the cortical surface at different scale levels with or without partial volume effect correction were evaluated on the [18F]FDG brain data before and after application of the EQ·PET filter. The EQ·PET methodology succeeded in finding the optimal smoothing that minimised root-mean-square error (RMSE) calculated using human brain [18F]FDG-PET datasets of EOAD patients, providing harmonized comparisons in the neurological context. Performance was superior for TOF than for TOF + PSF reconstructions. Results showed the capability of the EQ·PET methodology to minimize reconstruction-induced variabilities between brain [18F]FDG-PET images. However, moderate variabilities remained after harmonizing PSF reconstructions with standard non-PSF OSEM reconstructions, suggesting that precautions should be taken when using PSF modelling.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain / diagnostic imaging*
  • Fluorodeoxyglucose F18
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Image Processing, Computer-Assisted / standards
  • Phantoms, Imaging
  • Positron-Emission Tomography / methods*
  • Positron-Emission Tomography / standards
  • Radiopharmaceuticals

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18