Dynamic nuclear polarisation of liquids at one microtesla using circularly polarised RF with application to millimetre resolution MRI

J Magn Reson. 2019 Aug:305:138-145. doi: 10.1016/j.jmr.2019.06.013. Epub 2019 Jun 29.

Abstract

Magnetic resonance imaging in ultra-low fields is often limited by mediocre signal-to-noise ratio hindering a higher resolution. Overhauser dynamic nuclear polarisation (O-DNP) using nitroxide radicals has been an efficient solution for enhancing the thermal nuclear polarisation. However, the concurrence of positive and negative polarisation enhancements arises in ultra-low fields resulting in a significantly reduced net enhancement, making O-DNP far less attractive. Here, we address this issue by applying circularly polarised RF. O-DNP with circularly polarised RF renders a considerably improved enhancement factor of around 150,000 at 1.2 μT. A birdcage coil was adopted into an ultra-low field MRI system to generate the circularly polarised RF field homogeneously over a large volume. We acquired an MR image of a nitroxide radical solution with an average in-plane resolution of 1 mm. De-noising through compressive sensing further improved the image quality.

Keywords: Dynamic nuclear polarization; Low field; MRI; Nitroxide radical; Overhauser; Ultra low field.