Warning criteria for MEP monitoring during carotid endarterectomy: a retrospective study of 571 patients

J Clin Monit Comput. 2020 Jun;34(3):589-595. doi: 10.1007/s10877-019-00345-5. Epub 2019 Jul 2.

Abstract

Monitoring of transcranial electrical motor evoked potentials (tcMEP) during carotid endarterectomy (CEA) has been shown to effectively detect intraoperative cerebral ischemia. The unique purpose of this study was to evaluate changes of MEP amplitude (AMP), area under the curve (AUC) and signal morphology (MOR) as additional MEP warning criteria for clamping-associated ischemia during CEA. Therefore, the primary outcome was the number of MEP alerts (AMP, AUC and MOR) in the patients without postoperative motor deficit (false positives). We retrospectively reviewed data from 571 patients who received CEA under general anesthesia. Monitoring of somatosensory evoked potentials (SSEP) and tcMEP was performed in all cases (all-or-none MEP warning criteria). The percentages of false positives (primary parameter) of AMP, AUC and MOR were evaluated according to the postoperative motor outcome. In the cohort of 562 patients, we found significant SSEP/MEP changes in 56 patients (9.96%). In 44 cases (7.83%) a shunt was inserted. Nine patients (1.57%) were excluded due to MEP recording failure. False positives were registered for AMP, AUC and MOR changes in 121 (24.01%), 148 (29.36%) and 165 (32.74%) patients, respectively. In combination of AMP/AUC and AMP/AUC/MOR false positives were found in 9.52% and 9.33% of the patients. This study is the first to evaluate the correctness of the MEP warning criteria AMP, AUC and MOR with regard to false positive monitoring results in the context of CEA. All additional MEP warning criteria investigated produced an unacceptably high number of false positives and therefore may not be useful in carotid surgery for adequate detection of clamping-associated ischemia.

Keywords: Carotid endarterectomy; Intraoperative neurophysiologic monitoring; Motor evoked potentials; Warning criteria.

MeSH terms

  • Aged
  • Anesthesia, General / methods*
  • Anesthetics / pharmacology
  • Area Under Curve
  • Brain Ischemia / diagnostic imaging
  • Endarterectomy, Carotid / methods*
  • Evoked Potentials, Motor / physiology
  • Evoked Potentials, Somatosensory / physiology
  • False Positive Reactions
  • Female
  • Humans
  • Intraoperative Neurophysiological Monitoring / methods
  • Male
  • Middle Aged
  • Monitoring, Physiologic / methods*
  • Neurophysiology
  • Neurosurgical Procedures / methods
  • Retrospective Studies

Substances

  • Anesthetics