A Signal Acquisition Setup for Ultrashort Echo Time Imaging Operating in Parallel on Unmodified Clinical MRI Scanners Achieving an Acquisition Delay of [Formula: see text]

IEEE Trans Med Imaging. 2020 Jan;39(1):218-225. doi: 10.1109/TMI.2019.2924057. Epub 2019 Jun 20.

Abstract

Ultrashort echo time imaging on clinical systems is still limited by the rather long radio frequency switching times achievable with standard front end concepts. In this contribution, an independent parallel receive-only system is interfaced to an unmodified clinical MRI system, enabling imaging of species with ultrashort relaxation times, such as bone, tendon, teeth, or lung tissue. Synchronization of the system is achieved by an electronically decoupled one-way trigger line, a clock reference signal, and RF pulse tracking, thus ensuring minimal interference with the host system. With the proposed system, an acquisition delay of [Formula: see text] is experimentally demonstrated.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Femur / diagnostic imaging
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Signal Processing, Computer-Assisted*