We report two patients presenting with delayed complicationsafter Vibrant SounBridge® middle ear implant surgery: in both cases, a revision surgery was performed and lysis of the long process of the incus was highlighted. A re-assembly of the clip around the remaining long process of the incus was performed, associated with hydroxyapatite bone cement application, on the clip and the incudo-stapedial joint. Both patients had a satisfying result, with a mean follow up of 12 months.
Keywords: Hydroxyapatite bone cement; Long process of the incus; Revision; Vibrant soundbridge.
Copyright © 2019 Elsevier B.V. All rights reserved.