Measurement of the absolute number of photons of the hard X-ray beamline at the Linac Coherent Light Source

J Synchrotron Radiat. 2019 Mar 1;26(Pt 2):320-327. doi: 10.1107/S1600577519000250. Epub 2019 Feb 11.

Abstract

X-ray free-electron lasers provide intense pulses of coherent X-rays with a short pulse duration. These sources are chaotic by nature and therefore, to be used at their full potential, require that every X-ray pulse is characterized in terms of various relevant properties such as intensity, photon energy, position and timing. Diagnostics are for example installed on an X-ray beamline to specifically monitor the intensity of individual X-ray pulses. To date, these can however only provide a single-shot value of the relative number of photons per shot. Here are reported measurements made in January 2015 of the absolute number of photons in the hard X-ray regime at LCLS which is typically 3.5 × 1011 photons shot-1 between 6 and 9.5 keV at the X-ray Pump-Probe instrument. Moreover, an average transmission of ≉62% of the hard X-ray beamline over this energy range is measured and the third-harmonic content of ≉0.47% below 9 keV is characterized.

Keywords: FEL; absolute intensity; hard X-rays.