Influence of proper or reciprocating optimum torque reverse kinematics on cyclic fatigue of four single files

J Investig Clin Dent. 2019 Aug;10(3):e12409. doi: 10.1111/jicd.12409. Epub 2019 Mar 7.

Abstract

Aim: The aim of the present study was to evaluate the resistance to cyclic fatigue of heat-treated single files used in proper or in reciprocating optimum torque reverse motions.

Methods: Ninety-six HyFlex electrical discharge machining (EDM), Reciproc R25 (Rec), WaveOne Primary (WO), and WaveOne Gold (WOG) Primary were used. Resistance to cyclic fatigue was determined by recording time to fracture in a stainless steel artificial canal. Differential scanning calorimetric analysis was performed. Data were analyzed by two-way analysis of variance and Bonferroni post-hoc tests at 0.05.

Results: In both movements, HyFlex EDM showed higher time to fracture than other files (P < 0.0001). Rec showed higher time to fracture than WO (P < 0.0001), while WOG showed higher cyclic fatigue resistance than Rec and WO with both movements (P < 0.005).

Conclusions: Reciprocating optimum torque reverse motion improved significantly cyclic fatigue resistance of gold and controlled memory wire single files.

Keywords: differential scanning calorimetric analysis; fatigue resistance; nickel-titanium; optimum torque reverse; reciprocation.