Image quality and dose evaluation of MVCT TomoTherapy acquisitions: A phantom study

Phys Med. 2019 Jan:57:200-206. doi: 10.1016/j.ejmp.2019.01.009. Epub 2019 Jan 15.

Abstract

Background: The aim of this study is to evaluate the dose delivered and the image quality of pre-treatment MVCT images with Hi-Art TomoTherapy system, varying acquisition and reconstruction parameters.

Materials and methods: Catphan 500 MVCT images were acquired with all acquisition pitch and reconstruction intervals; image quality was evaluated in terms of noise, uniformity, contrast linearity, contrast-to-noise ratio (CNR) and spatial resolution with the Modulation Transfer Function (MTF). Dose was evaluated as Multi Slice Average Dose (MSADw) and measurements were performed with the Standard TomoTherapy® Quality Assurance Kit composed by the TomoTherapy Phantom, the Exradin A1SL ion chamber and TomoElectrometer. For each pitch-reconstruction interval, acquisitions were repeated 5 times.

Results: Differences in noise and uniformity, though statistically significant in some cases, were very small: noise ranged from 2.3% for Coarse - 3 mm to 2.4% for Coarse - 6 mm, while uniformity passed from 99.5% for Coarse - 6 mm to 99.8% for Normal - 4 mm. No differences at all were found for CNR for high and low density inserts, while MTF was higher for pitch Coarse, even if no differences in spatial resolution were observed visually (spatial resolution was up to 4 lp/cm for all combinations of pitch and reconstruction interval). Dose was dependent on pitch, being 1.0 cGy for Coarse, 1.5 cGy for Normal and 2.85 cGy for Fine.

Conclusions: We observed negligible differences in image quality among different pitch and reconstruction interval, thus, considerations regarding pre-treatment imaging modalities should be based only on dose delivered and on the desired resolution along the cranio-caudal axis for image-guided radiotherapy and adaptive radiotherapy purposes.

Keywords: Image guided radiotherapy; Image quality; Megavoltage computed tomography; Pre-treatment imaging; Tomotherapy.

MeSH terms

  • Phantoms, Imaging*
  • Quality Control
  • Radiation Dosage*
  • Radiotherapy Dosage
  • Radiotherapy, Image-Guided / instrumentation*
  • Tomography, X-Ray Computed*