Clinical and microbiological effects of dialyzers reuse in hemodialysis patients

J Bras Nefrol. 2019 Jul-Sep;41(3):384-392. doi: 10.1590/2175-8239-JBN-2018-0151. Epub 2019 Jan 24.
[Article in English, Portuguese]

Abstract

Introduction: Chronic kidney disease (CKD) has a high prevalence and is a worldwide public health problem. Reuse of dialyzers is a cost reduction strategy used in many countries. There is controversy over its effects on clinical parameters and microbiological safety.

Methods: In this clinical crossover study, 10 patients performed consecutive hemodialysis (HD) sessions divided in two phases: "single use" sessions (N = 10 HD sessions) followed by "dialyzer reuse" sessions (N = 30 HD sessions). Clinical, laboratory, and microbiological parameters were collected in the following time points: "single use", 1st, 6th, and 12th sessions with reuse of dialyzers, including bacterial cultures, endotoxins quantification in serum and dialyzer blood chamber, and detection of hemoglobin and protein residues in dialyzers.

Results: Mean age of the sample was 37 ± 16 years, 6 (60%) were men, and 5 (50%) were white. CKD and HD vintage were 169 ± 108 and 47 (23-111) months, respectively. Serum C-reactive protein (CRP) [4.9 (2.1) mg/mL], ferritin (454 ± 223 ng/mL), and endotoxin levels [0.76 (0.61-0.91) EU/mL] were high at baseline. Comparison of pre- and post-HD variations of serum levels of CRP and endotoxins in the "single use" versus "reuse" phases did not result in differences (p = 0.8 and 0.4, respectively). Samples of liquid in the dialyzer inner chamber were negative for the growth of bacteria or endotoxins. There was no significant clinical manifestation within and between the phases.

Conclusion: Dialyzers reuse was safe from a clinical, microbiological, and inflammatory point of view. The dialyzer performance remained adequate until the 12th reuse.

Publication types

  • Clinical Trial

MeSH terms

  • Adult
  • C-Reactive Protein / analysis
  • Cross-Over Studies
  • Endotoxins / blood
  • Equipment Reuse*
  • Female
  • Ferritins / blood
  • Follow-Up Studies
  • Humans
  • Inflammation / blood
  • Kidneys, Artificial / adverse effects*
  • Kidneys, Artificial / microbiology*
  • Male
  • Middle Aged
  • Pilot Projects
  • Renal Dialysis / instrumentation*
  • Renal Insufficiency, Chronic / therapy
  • Young Adult

Substances

  • Endotoxins
  • C-Reactive Protein
  • Ferritins